Feature :

  • Fully automatic, random access.
  • Flux
    • 600 Test / h for biochemical analysis.
    • 600 tests / h with ISE only (optional).
  • 90 reagent positions and 115 sample positions.
  • LIS interface, support remote maintenance.
  • The probe sample automatically detects blood clots, detects liquid level and collision protection function.
  • Functional tests for HbA1c.
  1. Equipment overview
    • Equipment type: Fully automatic, separate, STAT model priority
    • Speed :
      • 600 Test / h for biochemical analysis.
      • 600 tests / h with ISE only (optional)
    • Measuring Principle: Colorimetry, turbidimetry, ISE
    • Analytical method:
    • Endpoint, kinetics, fixed time ,…
    • Support single / double wavelength and 1-2 sample multi reagent, linear and non-linear standard.
    • Simultaneous analysis samples: 88 colorimetric samples and 3 ISE samples (optional K, Na, Cl)
  2. Sample location:
    • Sample position: 115 sample positions including: 50 routine sample positions, 20 STAT positions, 34 standard positions, 8 QC positions, 3 washing chemical positions. Continuously cool the standard position and QC position to keep the temperature from 5 ~ 15˚ C within 24 hours.
    • Sample Cuvette: Standard cup, original (blood) sample tube, different sizes (Ø12 ~ 16) mm × (25 ~ 100) mm.
    • Sample barcode: 12of5 、 code128 、 code39 、 UPC / EAN 、 code 93
    • Sample volume: 2µl-35µl, 0.1µl stepping.
    • Sampling technology: Detects liquid level, detects clogging and resists shock
    • Cross contamination rate: Automatically wash with hot water, carry rate ≤0.1%
    • Automatic sample phase: 3 ~ 170 times
  3. Reagent location
    • Suction needle (probe): 2, with liquid level detection and shock resistance function
    • Reagent volume: 20µl-350µl, 1 µl stepping.
    • Reagent location: 2 plates of reagent, 45×2, continuously cooled from 5 ~ 15˚C within 24h, containing 70ml, 20ml of reagent.
    • Reagent barcode: Code 128
    • Cross-contamination rate: Automatically wash with hot water, cross contamination rate ≤0.1%.
  4. Reaction location
    • Reaction Cuvet: 120 positions of plastic optical cuvette, optical diameter 6mm
    • Total reaction volume: 150µl-450µl
    • Reaction temperature: 37˚C, ± 0.1˚C
    • Stable temperature of the reaction plate: Circulating water
    • Mixing needle: 2, mix well after smoking reagent.
    • Wash reaction cuvette: 7 washing stations with 11 steps, cleaned with warm water.
    • Wastewater treatment:
      • There are two levels of flow with low and high concentration of wastewater.
      • There is a concentrated wastewater warning.
  5. Optical system
    • Light source: Halogen lamp: 20W / 12V. Lifespan is about 2000 hours
    • Monochromator : Grating photometry
    • Photoelectron road : Rear spectrophotometry
    • Wavelength: 340nm, 380nm, 405nm, 450nm, 480nm, 505nm, 546nm, 570nm, 600nm, 660nm, 700nm, 750nm or 800nm.
    • Detector: Bipolar LED
    • OD linear range: 0 ~ 3.3 Abs
  6. Standards and QC
    • Standard method: 1 point, 2 point, multipoint, and pointless linearity method.
    • QC method: real-time QC, one-by-one and monthly QC.
    • Handling samples out of control: The test piece is alerted when out of control, with the cause recorded.
  7. Operating system
    • Operating System: Windows XP
    • Analysis software: Using graphic software in English
    • Main functions of the software: Automatic calibration, automatic barcode scanning, test sample checking, management of reagent information, serum index, monitoring of the whole reaction process, preventing contamination. contamination, prevent possible cross-contamination, manage patient information, report and print results, ….
    • PC configuration :
      • CPU ≥ 2.2 GHz (dual-core processor), memory ≥ 1G, hard disk ≥ 160 G
      • 17 inch LCD display, laser printer, inkjet, inkjet (optional)
    • System connection: Connect via TCP / IP, RS-232C
  8. Other information
    • Dimensions: 1060mm x 790mm x 1150mm
    • Weight: 300 kg
    • Power supply: Voltage 220V ± 22V, 50Hz ± 1Hz, power source 2KVA.