Feature :
- Fully automatic, random access.
- Flux
- 600 Test / h for biochemical analysis.
- 600 tests / h with ISE only (optional).
- 90 reagent positions and 115 sample positions.
- LIS interface, support remote maintenance.
- The probe sample automatically detects blood clots, detects liquid level and collision protection function.
- Functional tests for HbA1c.
- Equipment overview
- Equipment type: Fully automatic, separate, STAT model priority
- Speed :
- 600 Test / h for biochemical analysis.
- 600 tests / h with ISE only (optional)
- Measuring Principle: Colorimetry, turbidimetry, ISE
- Analytical method:
- Endpoint, kinetics, fixed time ,…
- Support single / double wavelength and 1-2 sample multi reagent, linear and non-linear standard.
- Simultaneous analysis samples: 88 colorimetric samples and 3 ISE samples (optional K, Na, Cl)
- Sample location:
- Sample position: 115 sample positions including: 50 routine sample positions, 20 STAT positions, 34 standard positions, 8 QC positions, 3 washing chemical positions. Continuously cool the standard position and QC position to keep the temperature from 5 ~ 15˚ C within 24 hours.
- Sample Cuvette: Standard cup, original (blood) sample tube, different sizes (Ø12 ~ 16) mm × (25 ~ 100) mm.
- Sample barcode: 12of5 、 code128 、 code39 、 UPC / EAN 、 code 93
- Sample volume: 2µl-35µl, 0.1µl stepping.
- Sampling technology: Detects liquid level, detects clogging and resists shock
- Cross contamination rate: Automatically wash with hot water, carry rate ≤0.1%
- Automatic sample phase: 3 ~ 170 times
- Reagent location
- Suction needle (probe): 2, with liquid level detection and shock resistance function
- Reagent volume: 20µl-350µl, 1 µl stepping.
- Reagent location: 2 plates of reagent, 45×2, continuously cooled from 5 ~ 15˚C within 24h, containing 70ml, 20ml of reagent.
- Reagent barcode: Code 128
- Cross-contamination rate: Automatically wash with hot water, cross contamination rate ≤0.1%.
- Reaction location
- Reaction Cuvet: 120 positions of plastic optical cuvette, optical diameter 6mm
- Total reaction volume: 150µl-450µl
- Reaction temperature: 37˚C, ± 0.1˚C
- Stable temperature of the reaction plate: Circulating water
- Mixing needle: 2, mix well after smoking reagent.
- Wash reaction cuvette: 7 washing stations with 11 steps, cleaned with warm water.
- Wastewater treatment:
- There are two levels of flow with low and high concentration of wastewater.
- There is a concentrated wastewater warning.
- Optical system
- Light source: Halogen lamp: 20W / 12V. Lifespan is about 2000 hours
- Monochromator : Grating photometry
- Photoelectron road : Rear spectrophotometry
- Wavelength: 340nm, 380nm, 405nm, 450nm, 480nm, 505nm, 546nm, 570nm, 600nm, 660nm, 700nm, 750nm or 800nm.
- Detector: Bipolar LED
- OD linear range: 0 ~ 3.3 Abs
- Standards and QC
- Standard method: 1 point, 2 point, multipoint, and pointless linearity method.
- QC method: real-time QC, one-by-one and monthly QC.
- Handling samples out of control: The test piece is alerted when out of control, with the cause recorded.
- Operating system
- Operating System: Windows XP
- Analysis software: Using graphic software in English
- Main functions of the software: Automatic calibration, automatic barcode scanning, test sample checking, management of reagent information, serum index, monitoring of the whole reaction process, preventing contamination. contamination, prevent possible cross-contamination, manage patient information, report and print results, ….
- PC configuration :
- CPU ≥ 2.2 GHz (dual-core processor), memory ≥ 1G, hard disk ≥ 160 G
- 17 inch LCD display, laser printer, inkjet, inkjet (optional)
- System connection: Connect via TCP / IP, RS-232C
- Other information
- Dimensions: 1060mm x 790mm x 1150mm
- Weight: 300 kg
- Power supply: Voltage 220V ± 22V, 50Hz ± 1Hz, power source 2KVA.